Sauter TD 225-0.1US Käyttöohje

Sauter Mittauslaitteet TD 225-0.1US

Lue alta 📖 käyttöohje suomeksi merkille Sauter TD 225-0.1US (7 sivua) kategoriassa Mittauslaitteet. Tämä opas oli hyödyllinen 18 henkilölle ja sai 4.5 tähden keskimäärin 9.5 käyttäjältä

Sivu 1/7
Sauter GmbH
Ziegelei 1
D-72336 Balingen
e-mail: info@kern-sohn.com
Phone : +49-[0]7433- 9933-0
Fax: +49-[0]7433- - 9933 149
Internet: www.sauter.eu
Instruction Manual Ultrasonic Material
Thickness Gauge
SAUTER -TD US
Version 2.0
04/2020
GB
TD_US- -e-2020 BA
PROFESSIONAL MEASURING
2 TD_US- -e-2020 BA
GB
Congratulations on the purchase of an ultrasonic material thickness gauge from
SAUTER. We hope you will enjoy your quality measuring device with its wide range of
functions.
For questions, wishes or suggestions we are at your disposal.
Table of contents:
1. Functions ........................................................................................................ 3
2. Technical data ................................................................................................ 3
2.1 Scope of delivery: ..................................................................................................................... 3
3. Description of the control panel ................................................................... 4
4. Material selection ........................................................................................... 4
5. Calibration (Adjustment) ............................................................................... 5
6. Measuring procedure .................................................................................... 6
7. Measuring with adjustment of the ultrasonic speed ................................... 6
8. Battery replacement ...................................................................................... 7
TD_US- -e-2020 3 BA
Note: It is strongly recommended to calibrate the new instrument before rst
use This will result in a higher measurement accuracy from the beginning. .
1. Functions
The exclusive LSI microcomputer provides high measurement accuracy.
The device has a high emitting power and a wide sensitivity spectrum in terms of
sensitivity. It can recognize or assign sensors of dierent frequencies. It can also
measure rough surfaces such as cast iron. It is used in almost all branches of industry.
Suitable for measuring the material thickness of many materials, e.g. steel, cast iron,
aluminium, copper, brass, zinc, quartz glass, polyethylene, PVC, grey cast iron,
nodular cast iron.
Automatic shutdown function to maintain power.
Connectable to a PC for data transmission with RS232C cable and software, available
as optional accessory.
2. Technical data
TD- US
Display
4 digits, 10mm LCD display
Measuring range
1.2-225mm (45# steel)
Measurement
uncertainty
±0,5% n +0,1
Sound velocity
500-9000m/s
Ambient temperature
0-50°C
Humidity
≤80%
Power supply
4x 1.5V AAA (UM-4) batteries
Size
120x62x30mm
Weight
Approx. 164g (without batteries)
2.1 Scope of delivery:
Transport case
Operating instructions
Ultrasonic sensor
Optional accessories:
1. 232 Cable and software RS-