Festo SFAM-62-5000L-TG12-PNLK-PNVBA-M12 Käyttöohje

Festo Ei luokiteltu SFAM-62-5000L-TG12-PNLK-PNVBA-M12

Lue alta 📖 käyttöohje suomeksi merkille Festo SFAM-62-5000L-TG12-PNLK-PNVBA-M12 (60 sivua) kategoriassa Ei luokiteltu. Tämä opas oli hyödyllinen 15 henkilölle ja sai 4.5 tähden keskimäärin 2 käyttäjältä

Sivu 1/60
SFAM-...-PNLK
Flow sensor
Operating instruc-
tion
8196453
8196453
2024-03
[8196455]
Original instructions
IO-Link is a registered trademark of its respective trademark holder in certain countries.
Festo — SFAM-...-PNLK — 2024-03
3
Table of contents
1 Applicable documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 Safety instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Training of qualified personnel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Additional information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4 Product overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.1 Structure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.1.1 Product design. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.1.2 Display components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.2 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.2.1 Functional principle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.2.2 Operating statuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.2.3 Switching outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.2.4 Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2.5 Analogue output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.6 Security code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.7 Minimum value and maximum value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.8 Replicate parameter function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.9 Switching of standard conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.10 Zero point synchronisation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.1 Installation, pneumatic. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.1.1 Connecting the sensor with MS series service unit components. . . . . . . . . . . 19
5.2 Installation, electrical. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
6 Commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
6.1 Switching on the sensor in RUN mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
6.2 Displaying parameters in SHOW mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.3 Configuring the sensor in EDIT mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.3.1 Starting EDIT Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.3.2 Configuring the switching output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.3.3 Setting the volume pulse output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.3.4 Setting the analogue output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.3.5 Changing device settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.3.6 Replicating parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.3.7 Zero point synchronisation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.4 Teach-in switching points in TEACH mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.5 Measure volume or energy manually in RECORDER mode . . . . . . . . . . . . . . . . . . . . . . . . 27
4
Festo — SFAM-...-PNLK — 2024-03
7 IO-Link interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.1 General information, IO-Link. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.2 Identification parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.3 IO-Link default parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.4 IO-Link system commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.5 Smart sensor profile parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.6 Device-specific parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.7 IO-Link teach-in. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.8 Block parameterisation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.9 Process Data Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.10 IO-Link diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8.1 Reset sensor to factory setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
9 Fault clearance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
10 Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
11 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
12 Examples for calculating the maximum error of the display. . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Festo — SFAM-...-PNLK — 2024-03
Additional information
5
1 Applicable documents
All available documents for the product
èwww.festo.com/sp.
2 Safety
2.1 Safety instructions
Only use the product in its original condition without unauthorised modifications.
Only use the product if it is in perfect technical condition.
Observe the identifications on the product.
Condensation, oil mist, foreign matter and other contaminants in the compressed air can damage
the product. Use media only in accordance with the specifications
è11 Technical data.
Take into consideration the ambient conditions at the location of use.
The product may generate high frequency interference, which may require interference suppression
measures in residential areas.
2.2 Intended use
The flow sensor monitors the flow rate, volume, pressure and temperature of gaseous media in piping
systems or terminals in industry.
2.3 Training of qualified personnel
Work on the product may only be carried out by qualified personnel who can evaluate the work and
detect dangers. The qualified personnel have skills and experience in dealing with electropneumatic
(open-loop) control technology.
3 Additional information
Contact the regional Festo contact if you have technical problems
èwww.festo.com.
Accessories èwww.festo.com/catalogue.
6
Festo — SFAM-...-PNLK — 2024-03
Product overview
4 Product overview
4.1 Structure
4.1.1 Product design
1
2 3 4 5
6
7
Fig. 1: SFAM-62-...-T...-PNLK-PNVBA-M12
1
Pneumatic port 1
2
Laminar flow inlet
3
Plug for electrical connection
4
[A] key
5
[Edit] key
6
[B] key
7
Pneumatic port 2
Festo — SFAM-...-PNLK — 2024-03
Product overview
7
1
2 3 4 5
6
7
Fig. 2: SFAM-90-...-T...-PNLK-PNVBA-M12
1
Pneumatic port 1
2
Laminar flow inlet
3
Plug for M12 electrical connection
4
[A] key
5
[Edit] key
6
[B] key
7
Pneumatic port 2
8
Festo — SFAM-...-PNLK — 2024-03
Product overview
2
1
3
5
678
4
Fig. 3: SFAM-62-...-M...-PNLK-PNVBA-M12
1
Laminar flow cartridge
2
Pneumatic port 1
3
Display
4
Pneumatic port 2
5
[B] key
6
[Edit] key
7
[A] key
8
Plug for M12 electrical connection
1
23
5
678
4
Fig. 4: SFAM-90-...-M...-PNLK-PNVBA-M12
1
Laminar flow cartridge
2
Pneumatic port 1
3
Display
4
Pneumatic port 2
5
[B] key
6
[Edit] key
7
[A] key
8
Plug for M12 electrical connection
Festo — SFAM-...-PNLK — 2024-03
Product overview
9
4.1.2 Display components
1
2 2
3
4
56
Fig. 5: LCD display
1
Flow rate bar display
2
Output display
3
Main display
4
Bar displays for volume measurement,
pneumatic energy measurement or pulse
measurement
5
Secondary display
6
Status indicator
Symbols on the display
Symbols Description
‘ OutA’ Switching output OutA selected, flashes when IO-Link is active.
‘ OutA’ Switching output OutA set.
‘OutC’ Volume pulse OutC selected, flashes when IO-Link is active.
‘OutC’ Volume pulse OutC set.
‘OutB’ Switching output OutB selected.
‘OutB’ Switching output OutB set.
‘OutD’ Pressure measurement OutD selected.
‘OutD’ Pressure measurement OutD set.
Anlg’ Analogue output selected.
‘Pulse’ Volume pulse selected.
10
Festo — SFAM-...-PNLK — 2024-03
Product overview
Symbols Description
Information/input display
Graphic display of flow measurement
Graphic display of volume measurement, energy measurement or
pulse measurement
q’ Input signal flow rate selected.
‘p Input signal pressure selected.
‘t’ Input signal temperature selected.
‘Lock’ Security code activated.
‘Spec’ Special menu selected.
‘Option’ Parameters that influence the measured value display have been
changed from the factory setting: reference condition (standard) or
gas.
Tab. 1: Symbols on the display
Example for LCD display Meaning
Main display Secondary display
Measured value indicator and unit in RUN mode
‘1730’ ‘l/min’ Measured value indicator and unit
Menu for the switching outputs OutA, OutB, OutC and OutD
‘Edit’ ‘Flow’/‘bin Edit menu for the switching outputs
‘OFF’ ‘Fctn’ Switching function deactivated.
‘_I¯’ ‘Fctn’ Threshold value comparator
‘_I¯I_ ‘Fctn’ Window comparator
‘_[]¯ ‘Fctn’ 2-point threshold value comparator
‘1800’ ‘SP’ Switching point value
‘800’ ‘SP.Lo Value of lower switching point
‘1220’ ‘SP.Hi’ Value of upper switching point
‘12.0’ ‘HY’ Hysteresis value
‘NO’ ‘logic’ Switching characteristics of the switching outputs:
‘NO’ = N/O contact, normally open
‘NC’ = N/C contact, normally closed
Festo — SFAM-...-PNLK — 2024-03
Product overview
11
Example for LCD display Meaning
Main display Secondary display
‘bLUE’ ‘COLR’ Display colour
‘200’ ‘PULS’ / ‘MSEC’ Width of the volume pulse or mass pulse
Menu for the analogue output
‘Edit’ ANLG’/‘Out’ Edit menu for the analogue outputs
‘1._.5 V’ / ‘Out’ Analogue output type ‘0_10V’, ‘1_5V’, ‘4_20MA’
‘38 ‘In.Hi’/‘%’ Scaling of the analogue output: end value
‘3’ ‘In.Lo’/‘%’ Scaling of the analogue output: start value
Extreme values and average values in SHOW mode
‘510’ ‘MIN’/‘l/Min’ Minimum measured flow rate since switch-on or the last
reset
‘3080’ ‘MAX’/‘l/Min’ Maximum measured flow rate since switch-on or the last
reset
‘2080’ AVER’/‘l/Min’ Average of the flow measurement
Menu for device settings (Spec menu)
‘Edit’ ‘MENU’ Edit menu for device settings
‘64’ ‘Filt’/‘MSEC’ Value of the filter time constant for the flow rate measure-
ment signal
‘l/min’ ‘FLOW’/‘Unit’ Display unit for flow rate measurement
‘0°C’ ‘REF’/‘Cond’ Reference standard for gas volume
Air’ ‘GAS’ Selection of the operating medium
‘OFF’ ‘Z.Adj’ Activation of the zero point synchronisation
‘bar’ ‘PRSR’/‘Unit’ Display unit for pressure measurement
‘°C’ ‘tEMP’/‘Unit’ Display unit for temperature measurement
‘Unit’ ‘Sub.d’ Settings of the secondary display in RUN mode
‘10 ‘Eco’/‘SEC’ Economy mode: time after which the display background
lighting is switched off.
‘PNP’ ‘bin’/‘Out’ Switchover of the switching outputs between PNP and NPN
‘FLOW’ ‘bin’/‘Pin4’ Switchover of the switching signal at pin 4 between flow
monitoring and volume pulse
‘FLOW’ ‘bin’/‘Pin2’ Switchover of the signal at Pin 2 between flow monitoring,
flow measurement and temperature recording
12
Festo — SFAM-...-PNLK — 2024-03
Product overview
Example for LCD display Meaning
Main display Secondary display
‘OFF’ ‘Code’
‘Lock’ flashes
Activation and specification of the security code
‘OFF’ ‘MASt’ Activation of the IO-Link master function for replication of
parameters
Tab. 2: Example for LCD display
4.2 Function
4.2.1 Functional principle
The sensor measures the standard volumetric flow rate or mass flow rate by a thermal process. The
sensor measures the amount of heat extracted from the heated surface of a micro-mechanical sensor
element by the medium as it flows past. The sensor calculates the flow rate or the cumulative volume
from the extracted heat and shows the result on the display. The sensor calculates the temperature of
the medium from the sensor signals.
The sensor measures the relative pressure with a piezoresistive sensor element at the pneumatic
output. The temperature measurement value is calculated from the signals from the thermal meas-
uring element.
The sensor can be connected to higher-level systems via switching outputs, a volume switching
pulse, an analogue output or an IO-Link interface.
The switching outputs monitor a threshold value or a signal range depending on the configuration.
The switching outputs can be set as PNP or NPN and N/O contact (NO) or N/C contact (NC).
The analogue output can be configured either as a voltage output or as a current output.
The process values can be read out and parameters changed and transmitted to additional devices
through the IO-Link interface.
If the flow rate is included, a volume signal or a mass signal can be calculated and output by a pulse
output and also a cumulative measured value via IO-Link.
Festo — SFAM-...-PNLK — 2024-03
Product overview
13
Fig. 6: SFAM signal structure
4.2.2 Operating statuses
Operating status Function
RUN mode Basic state after the operating voltage has been applied.
Display of the current measured value.
Display of the selected inputs and outputs.
Switch between the flow rate, pressure, pneumatic power, volume/mass
and temperature measured variables.
SHOW mode Display of the current settings for the switching outputs and the ana-
logue output.
Display and reset of the minimum and maximum values.
Display of the average flow rate measurement.
EDIT mode Setting or modifying parameters.
TEACH mode Acceptance of the current measured value to specify switching points.
RECORDER mode Manual measurement of volume, mass or pneumatic energy.
Tab. 3: Operating statuses
14
Festo — SFAM-...-PNLK — 2024-03
Product overview
4.2.3 Switching outputs
4.2.3.1 Switching functions
Threshold value comparator
Function Normally open contact (N/O) Normally closed contact (N/C)
Switching function:
1 switching point (SP)
TEACH mode:
2 teach points (TP1, TP2)
SP = ½ (TP1 + TP2)
Tab. 4: Threshold value comparator
Window comparator
Function Normally open contact (N/O) Normally closed contact (N/C)
Switching function:
2 switching points (SP.Lo,
SP.Hi)
TEACH mode:1)
2 teach points (TP1, TP2)
TP1 = SP.Lo, TP2 = SP.Hi
1) SP.Lo = lower value, SP.Hi = higher value, independent of the teach sequence
Tab. 5: Window comparator
2-point threshold value comparator
Function N/O (normally open) N/C (normally closed)
Switching function:
2 switching points (SP.Lo,
SP.Hi)
TEACH mode:1)
2 teach points (TP1, TP2)
TP1 = SP.Lo, TP2 = SP.Hi
1) SP.Lo = lower value, SP.Hi = higher value, independent of the teach sequence
Tab. 6: 2-point threshold value comparator
4.2.3.2 Colour change
A red colour change can be set in the display for the OutA, OutB and OutD switching outputs
depending on the switching status. The colour change enables the system status to be identified from
a greater distance. The red display is always dominant.
Festo — SFAM-...-PNLK — 2024-03
Product overview
15
Parameters Meaning
‘bLUE’ The display is always blue; the colour change function is switched off.
‘R.On’ The display is red when the switching output is set (high = 1).
The display is blue when the switching output is not set (low = 0).
‘R.OFF’ The display is red when the switching output is not set (low = 0).
The display is blue when the switching output is set (high = 1).
Tab. 7: Change in colour of display
4.2.3.3 Volume measurement and mass measurement
A threshold value for the volume or the mass is set in the accumulated volume measurement or
mass measurement. If the configured threshold value is reached, the sensor outputs a switching
pulse at switching output OutC. The volume measurement or mass measurement is restarted at every
switching pulse. The pulse width is adjustable.
Setting of N/O contact (NO) Setting N/C contact (NC)
Tab. 8: Volume pulse with accumulated volume measurement
4.2.4 Filter
The signal filter smoothes the flow rate measurement signal and the pressure measurement signal.
Smoothing affects the outputs as shown in the following figures. The switching times of the switching
outputs along with the rise time and fall time of the analogue output change. The filter time equals the
time constant of a low-pass filter.τ
Fig. 7: Effect of filter settings on the flow rate measurement signal
Festo — SFAM-...-PNLK — 2024-03
Product overview
17
The minimum and maximum values are reset when the operating voltage is switched off.
4.2.8 Replicate parameter function
This function enables all settings that have been carried out on one sensor (master) to be transferred
to other identical sensors (device).
Parameters are transferred with the IO-Link functions. The configured sensor is set to a master mode
and can then send its parameters to an identical device sensor (identical device ID).
4.2.9 Switching of standard conditions
The flow rate units and volume units that can be displayed can be based on different standard
conditions. It is possible to switch between the following standard conditions. The corresponding
standard is defined by selection of the reference temperature in the menu guiding on the display.
REF/Cond Off 15°C 20°C
Standard DIN 1343 ISO 2533 ISO6358/
ISO8778
Air pressure (absolute) [kPa] 101.325 101.325 100
Temperature [°C] 0 15 20
Humidity [%] 0 0 65
Status information "Option" Light off Light on Light on
Correction factor for measurement range
end value
1 1.055 1.087
Tab. 9: Standard conditions for flow rate units and volume units
The sensor is factory-calibrated to the physical standard conditions in accordance with DIN1343.
If a standard other than DIN1343 is selected, the specified measurement range (±100%FS)changes
in value by the factor specified in
èTab. 9 Standard conditions for flow rate units and volume units.
This change is visualised in the display by ‘Option’.
The measured value on the sensor is adjusted by switching the reference standard. When using the
analogue output in combination with another reference standard, the effect on the nominal measuring
range end value must be taken into account according to the correction factor
èTab. 9 Standard
conditions for flow rate units and volume units.
4.2.10 Zero point synchronisation
If the sensor is installed in a vertical position, the measurement may deviate at high pressures. The
zero point can be adjusted to compensate for the measurement deviation. The zero point adjustment
affects the display and all outputs. The zero point adjustment is possible only for the Zero Adjust = ON
setting and the zero range hiding is no longer active. The offset error is visible on the display and can
be set to zero using a zero-adjust teach process.
18
Festo — SFAM-...-PNLK — 2024-03
Installation
5 Installation
5.1 Installation, pneumatic
NOTICE
An unfavourable mounting position may impair the function of the product.
When selecting the mounting position, make sure that condensate from the compressed air lines
cannot accumulate in the sensor.
Mounting position:
Horizontal ±5°
Vertical with zero point adjustment è4.2.10 Zero point synchronisation
The air mass flow is routed to the port at which the laminar flow inlet with the SFAM-…-T/-W... or
laminar flow cartridge with the SFAM-…-M... is located. The air mass flow is taken from the opposite
port.
Fig. 10: SFAM-...-T... with laminar flow inlet
20
Festo — SFAM-...-PNLK — 2024-03
Commissioning
5.2 Installation, electrical
WARNING
Risk of injury due to electric shock.
For the electrical power supply, use only PELV circuits in accordance with IEC 60204-1/EN 60204-1
(Protective Extra-Low Voltage, PELV).
Observe the general requirements of IEC 60204-1/EN 60204-1 for PELV circuits.
Only use voltage sources that ensure a reliable electric separation from the mains network in
accordance with IEC 60204-1/EN 60204-1.
1. Please note the maximum cable length for the technical connection
è11 Technical data.
2. Configure binary outputs according to the wiring.
Tightening torque for the union nut at the plug: max. 0.3Nm
Plug Pin Wire colour1) Assignment
M12, 5-pin 1 Brown (BN) Operating voltage +24VDC
2 White (WH) Electrical output 2 (OutB, OutD or Analog)
3 Blue (BU) Operating voltage 0V
4 Black (BK) Electrical output 1 (OutA, OutC, C/Q line at IO-
Link)
5 Grey (GY) N/C
1) When using the connecting cable from the accessories.
Tab. 11: Pin allocation
Circuit diagram
Fig. 12: Circuit diagram
6 Commissioning
6.1 Switching on the sensor in RUN mode
Switch on the operating voltage.
ÄThe current measured value is displayed. The sensor is in the basic status.
The sensor is in the basic status in RUN mode. The sensor shows the current measured values. The
basic status can be reached as follows from other modes:
Press and hold the [Edit] key for 3seconds.
After expiration of a monitoring period (timeout).
Festo — SFAM-...-PNLK — 2024-03
Commissioning
21
Switchover of measured value indicator
The measured value indicator can be switched between flow rate, pressure temperature, volume/mass
and power in RUN mode.
Requirement: the sensor is ready for operation and is in RUN mode.
1. Briefly press the [A] or [B] key.
2. Repeat the process with a 1…3 second pause until the desired measured variable is displayed.
6.2 Displaying parameters in SHOW mode
Requirement: the sensor is ready for operation and is in RUN mode.
RUN mode is reached in SHOW mode as follows:
Press and hold the [Edit] key for 3 seconds.
Press and hold the [A] or [B] key 3 seconds.
Switching output OutA or switching output OutC
1. Press the [A] key twice in rapid succession.
ÄThe first set parameter is displayed.
2.
Press the [B] key to display the preceding parameters.
Press the [A] key to display the following parameters
èFig. 13.
ÄThe flow rate value ‘AVER’ is displayed last.
Switching output OutB, switching output OutD or analogue output Anlg
1. Press the [B] pushbutton twice in rapid succession.
ÄThe first set parameter is displayed.
2.
Press the [A] key to display the preceding parameters.
Press the [B] key to display the following parameters
èFig. 13.
ÄThe flow rate value ‘MAX’ is displayed last.
22
Festo — SFAM-...-PNLK — 2024-03
Commissioning
Button [A] or button [B]
Double click button [A] or button [B]
Fig. 13: Menu structure for SHOW mode
Festo — SFAM-...-PNLK — 2024-03
Commissioning
23
6.3 Configuring the sensor in EDIT mode
Button [Edit]
Button [A] or button [B]
Default setting
Fig. 14: EDIT mode menu structure
Changes to the switching behaviour are effective immediately.
RUN mode is reached in EDIT mode as follows:
Press and hold the [Edit] key for 3 seconds.
24
Festo — SFAM-...-PNLK — 2024-03
Commissioning
6.3.1 Starting EDIT Mode
Requirement: the sensor is ready for operation and is in RUN mode.
1. Press the [Edit] key.
ÄThe EDIT mode is active. ‘OutA’ and ‘Edit’ are displayed.
2. If the security code ‘LOCK’ is active, enter the security code using the [A] pushbutton or the [B]
pushbutton and press the [Edit] key.
ÄThe EDIT mode is active. ‘OutA’ and ‘Edit’ are displayed.
3. Press the [A] or [B] key to switch between OutA, OutB, OutC, OutD, Anlg and Spec.
6.3.2 Configuring the switching output
The process is basically the same for configuring the switching outputs for OutA, OutB and OutD. In
the following, the process is described using the OutA switching output.
Requirement: EDIT mode is active.
1. Press the [Edit] key.
Ä‘Fctn’ flashes.
2. Select the switching function ‘OFF’, ‘_I¯’, ‘_I¯I_’ or ‘_[]¯’ with the [A] or [B] key.
3. Press the [Edit] key.
ÄThe set parameter value is saved. The next adjustable parameter is shown.
4. Set the parameter value with the [A] or the [B] key.
5. Repeat steps 3 and 4 until all parameters are set èFig. 14.
6. Press the [Edit] key.
ÄRUN mode is active.
6.3.3 Setting the volume pulse output
Requirement: EDIT mode is active.
1. Select ‘VOL’/‘PULS’ with the [A] or the [B] key.
2. Press the [Edit] key.
Ä‘SP’ and the set volume unit are displayed alternately.
3. Set the parameter value with the [A] or the [B] key.
4. Press the [Edit] key.
ÄThe set parameter value is saved. The next adjustable parameter is shown.
5. Repeat steps 3 and 4 until all parameters are set.
6. Press the [Edit] key.
ÄRUN mode is active.
6.3.4 Setting the analogue output
Requirement: EDIT mode is active.
1. Select ‘ANLG’/‘Out’ with the A or the B key.
2. Press the [Edit] key.
Ä‘Out’/‘V’ or ‘Out’/‘MA’ are displayed alternately.
3. Select the parameter with the [A] or the [B] key.
Festo — SFAM-...-PNLK — 2024-03
Commissioning
25
4. Press the [Edit] key.
ÄThe set parameter value is saved. The next adjustable parameter is shown.
5. Repeat steps 3 and 4 until all parameters are set.
6. Press the [Edit] key.
ÄRUN mode is active.
6.3.5 Changing device settings
Requirement: EDIT mode is active.
1. Select the ‘Spec’ with the [A] or the [B] key.
Ä‘Spec’ is displayed.
2. Press the [Edit] key.
Ä‘Filt’/‘MSEC’ are displayed alternately.
3. Set the parameter with the [A] or the [B] key.
4. Press the [Edit] key.
ÄThe set parameter value is saved. The next adjustable parameter is shown.
5. Repeat steps 3 and 4 until all parameters are set.
6. Press the [Edit] key.
ÄRUN mode is active.
6.3.6 Replicating parameters
Requirements:
The previously configured sensor (master sensor) is ready for operation and is in RUN mode.
The master sensor and device sensor are identical with reference to the parameters, i.e.they have
the same device ID.
The master sensor is connected to the device sensor and the power supply.
Parameterisation of the device sensor must not be blocked via IO-Link.
Master
Sensor
Power Suppply
Device
Sensor
Fig. 15: Replicate pin allocation parameters
1. Select the ‘SPEC’ special menu from the device settings on the master sensor.
2. Press the [Edit] key repeatedly until ‘MASt’ appears.
3. Select ‘ON’ with the [A] or the [B] key.
4. Press the [Edit] key.
Ä‘REPL’ and ‘REdY’ appear.
5. Press the [A] or the [B] key.
Ä‘REPL’ and ‘RUN’ appear briefly. The parameters are transmitted to the device sensor. ‘REPL’
and ‘REdY’ appear. If an error occurs, an error message appears
è9 Fault clearance.
26
Festo — SFAM-...-PNLK — 2024-03
Commissioning
Ä
6. Repeat point 5 if an additional sensor is to be parameterised.
7. Press the [Edit] key.
ÄRUN mode is active.
6.3.7 Zero point synchronisation
Requirements:
‘Z.Adj’ ‘ON’ is set.
The operating pressure is present, but there is no flow rate.
RUN mode is active.
The measured value lies in the range 0 l/min ± 3 %FS.
1. Press the [A] key and the [B] key.
2. Also press the [Edit] key.
If ‘OK’ appears: the zero point synchronisation was successful.
‘FAIL’ appears: the zero point synchronisation was not successful. Check requirements.
If [Z.AdJ] [OFF] is set for a later time, the sensor takes over the factory setting calibration values.
6.4 Teach-in switching points in TEACH mode
The switching points for flow monitoring can be established in the TEACH mode and the pressure
monitoring can be defined è4.2.3 Switching outputs.
Before the teach-in, set the switching function to EDIT mode è6.3 Configuring the sensor in EDIT
mode.
The process for teach-in of switching signal OutA, switching signal OutB and switching signal OutD is
basically the same:
OutA: press the [A] and the [Edit] key.
OutB or OutD: press the [B] key and the [Edit] key.
Sequence for the switching signal OutA
Requirements:
The sensor is ready for operation and is in RUN mode.
OutA must be selected at pin 4.
1. Generate the flow rate for TP1.
2. Press the [A] and the [Edit] key.
ÄWith security code activated: ‘Lock’ flashes.
3. Set the security code with the [A] or the [B] key.
4. Press the [Edit] key.
Ä‘t-IN’ flashes.
ÄThe flow rate measured value will be applied as teach point (TP1).
5. Generate the flow rate for TP2.
Festo — SFAM-...-PNLK — 2024-03
Commissioning
27
6. Press the [A] and the [Edit] key.
ÄThe flow rate measured value will be applied as teach point (TP2).
ÄThe ‘SP’ switching point of the ‘SP.Lo’ and ‘SP.Hi’ switching points are valid.
ÄThe RUN mode is displayed.
Sequence for the switching signal OutD
Requirements:
The sensor is ready for operation and is in RUN mode.
OutD must be selected at pin 2.
1. Apply the pressure value TP1.
2. Press the [B] and the [Edit] key.
ÄWith security code activated: ‘Lock’ flashes.
3. Set the security code with the [A] or the [B] key.
4. Press the [Edit] key.
Ä‘t-IN’ flashes.
ÄThe pressure measured value will be applied as the first teach point (TP1).
5. Apply the pressure value TP2.
6. Press the [B] and the [Edit] key.
ÄThe pressure measured value will be applied as the second teach point (TP2).
ÄThe ‘SP’ switching point of the ‘SP.Lo’ and ‘SP.Hi’ switching points are valid.
ÄThe RUN mode is displayed.
6.5 Measure volume or energy manually in RECORDER mode
Requirement: the sensor is ready for operation and is in RUN mode.
1. Press the [A] and [B] keys simultaneously.
ÄThe sensor is in RECORDER mode
ÄThe current measured value of the volume measurement or energy measurement is dis-
played.
The [Edit] key can be used to switch between volume measurement and energy measurement.
2. Press the [A] key to start or stop the volume measurement or the energy measurement.
ÄDuring startup the volume measurement and the energy measurement are started at the
same time.
ÄThe bar display for the volume recorder or the energy recorder runs from bottom to top.
ÄIf the volume measurement or the energy measurement is stopped, the bar display stops at a
point.
3. Press the [B] key to reset the volume measurement and the energy measurement to 0.
4. To exit RECORDER mode, press the [A] and [B] key simultaneously.
ÄThe RUN mode is displayed.
28
Festo — SFAM-...-PNLK — 2024-03
IO-Link interface
If the RECORDER mode is exited during a volume measurement or energy measurement, the volume
measurement or energy measurement continues in the background.
When the operating voltage is switched off, the volume measurement or energy measurement is saved
with a degree of inaccuracy, because the value is saved to the non-volatile memory area every 2
minutes.
7 IO-Link interface
7.1 General information, IO-Link
Characteristics Specification
Protocol IO-Link
Revision ID V1.1
Device profiles Smart Sensor - SSP 4.1.3
Firmware Update
identification and diagnosis
Function Quantity detection
Function Locator
Function Product URI
Transmission rate COM3
SIO-Mode support Yes
Port type class A
Process data output length 0 bit
Process data input length 96 bit
Process data content IN Flow measurement value 16 bit MDC
Flow monitoring 2 bit SSC
Volume pulse/mass pulse 1 bit SSC
Temperature measurement value 16 bit MDC
Temperature monitoring 2 bit SSC
Pressure measurement value 16 bit MDC
Pressure monitoring 2 bit SSC
Service data IN Volume measurement/mass measurement value 32 bit
Pneumatic energy measurement value 32 bit
Pneumatic power measurement value 32 bit
Min. cycle time 1,5ms
Festo — SFAM-...-PNLK — 2024-03
IO-Link interface
29
Characteristics Specification
Data storage required 0,5kByte
Vendor ID 0x014D (333)
Device ID èTab. 13 Device ID values
Tab. 12: General IO-Link specification
Device ID
[dec]
Device ID
[hex]
Order Code
329424 0x0506D0 SFAM-1000-PNVBA
329425 0x0506D1 SFAM-3000-PNVBA
329426 0x0506D2 SFAM-5000-PNVBA
329427 0x0506D3 SFAM-10000-PNVBA
329428 0x0506D4 SFAM-15000-PNVBA
Tab. 13: Device ID values
7.2 Identification parameters
Index Sub-
index
Access
1)
Name Value Format
0x0010
(16)
0x00 R Vendor Name Festo StringT (5Bytes)
0x0011
(17)
0x00 R Vendor Text www.festo.com StringT (20Bytes)
0x0012
(18)
0x00 R Product Name Order Code, e.
g. SFAM-62-1000L-M-
PNLK-PNVBA-M12
StringT
(max.64Bytes)
0x0013
(19)
0x00 R Product ID e. g. 563796 SFAM-1000-
PNVBA
StringT
(max.64Bytes)
0x0014
(20)
0x00 R Product Text Flow sensor StringT (11Bytes)
0x0015
(21)
0x00 R Serial Number Product Key, e. g.
KB16TVPZCBV
StringT (11Bytes)
0x0016
(22)
0x00 R Hardware Revision e. g. REV01 StringT (5Bytes)
0x0017
(23)
0x00 R Firmware Revision e. g. V8.8.12 StringT
(max.8Bytes)

Tuotetiedot

Merkki: Festo
Kategoria: Ei luokiteltu
Malli: SFAM-62-5000L-TG12-PNLK-PNVBA-M12

Tarvitsetko apua?

Jos tarvitset apua merkille Festo SFAM-62-5000L-TG12-PNLK-PNVBA-M12 esitä kysymys alla ja muut käyttäjät vastaavat sinulle




Ei luokiteltu Festo Käyttöohjeet

Ei luokiteltu Käyttöohjeet

Viimeisimmät Ei luokiteltu Käyttöohjeet